St. jude medical and carboMedics mechanical heart valves in the aortic position comparison of long-term results

We designed this study to compare long-term results of St. Jude Medical and CarboMedics mechanical heart valves in the aortic position. We retrospectively analyzed the results of 174 consecutive patients who received either a St. Jude (n=80) or a CarboMedics (n=94) mechanical aortic valve from March 1992 through October 2004. The follow-up rate was 97.7%. The mean follow-up duration for the St. Jude group was 79.3 +/- 35.0 and for the CarboMedics group, 70.0 +/- 34.3 months. The cumulative follow-up was 523.8 and 530.1 patient-years, respectively The 30-day mortality rates for the St. Jude and CarboMedics patients were 1.3% and 3.2%, respectively The actuarial survival rate for the St. Jude group at 138.0 +/- 4.7 months was 75.9% +/- 0.1% and for the CarboMedics group at 130.8 +/- 4.8 months was 69.8% +/- 0.1% (P=NS). There was no structural valve deterioration in either group. Freedom from thromboembolic events was 87.7% for the St. Jude group and 83.0% for the CarboMedics group (P = NS). Freedom from bleeding events for the St. Jude group was 93.6% and for the CarboMedics group, 89.7% (P=NS). The results obtained from this study indicate that standard St. Jude Medical and CarboMedics aortic valve prostheses offer similar excellent clinical performance. Definitive judgment must await trials that are extensive, randomized, and prospective.

Dergi Adı TEXAS HEART INSTITUTE JOURNAL
Dergi Cilt Bilgisi 33
Dergi Sayısı 2
Sayfalar 154 - 159
Yayın Yılı 2006
Eser Adı
[dc.title]
St. jude medical and carboMedics mechanical heart valves in the aortic position comparison of long-term results
Yazar
[dc.contributor.author]
Kandemir, Ozer
Yazar
[dc.contributor.author]
Tokmakoglu, Hilmi
Yazar
[dc.contributor.author]
Yildiz, Ulku
Yazar
[dc.contributor.author]
Tezcaner, Tevfik
Yazar
[dc.contributor.author]
Yorgancioglu, A. Cem
Yazar
[dc.contributor.author]
Gunay, Iihan
Yazar
[dc.contributor.author]
Suzer, Kaya
Yayın Yılı
[dc.date.issued]
2006
Yayıncı
[dc.publisher]
TEXAS HEART INST
Yayın Türü
[dc.type]
article
Özet
[dc.description.abstract]
We designed this study to compare long-term results of St. Jude Medical and CarboMedics mechanical heart valves in the aortic position. We retrospectively analyzed the results of 174 consecutive patients who received either a St. Jude (n=80) or a CarboMedics (n=94) mechanical aortic valve from March 1992 through October 2004. The follow-up rate was 97.7%. The mean follow-up duration for the St. Jude group was 79.3 +/- 35.0 and for the CarboMedics group, 70.0 +/- 34.3 months. The cumulative follow-up was 523.8 and 530.1 patient-years, respectively The 30-day mortality rates for the St. Jude and CarboMedics patients were 1.3% and 3.2%, respectively The actuarial survival rate for the St. Jude group at 138.0 +/- 4.7 months was 75.9% +/- 0.1% and for the CarboMedics group at 130.8 +/- 4.8 months was 69.8% +/- 0.1% (P=NS). There was no structural valve deterioration in either group. Freedom from thromboembolic events was 87.7% for the St. Jude group and 83.0% for the CarboMedics group (P = NS). Freedom from bleeding events for the St. Jude group was 93.6% and for the CarboMedics group, 89.7% (P=NS). The results obtained from this study indicate that standard St. Jude Medical and CarboMedics aortic valve prostheses offer similar excellent clinical performance. Definitive judgment must await trials that are extensive, randomized, and prospective.
Açıklama
[dc.description]
WOS: 000239032400011
Açıklama
[dc.description]
PubMed: 16878617
Kayıt Giriş Tarihi
[dc.date.accessioned]
2019-12-23
Açık Erişim Tarihi
[dc.date.available]
2019-12-23
Yayın Dili
[dc.language.iso]
eng
Konu Başlıkları
[dc.subject]
aortic valve
Konu Başlıkları
[dc.subject]
coronary thrombosis
Konu Başlıkları
[dc.subject]
heart valve prosthesis/adverse effects
Konu Başlıkları
[dc.subject]
postoperative complications
Konu Başlıkları
[dc.subject]
prosthesis design
Konu Başlıkları
[dc.subject]
thromboembolism/ prevention & control
Konu Başlıkları
[dc.subject]
treatment outcome
Haklar
[dc.rights]
info:eu-repo/semantics/closedAccess
ISSN
[dc.identifier.issn]
0730-2347
İlk Sayfa Sayısı
[dc.identifier.startpage]
154
Son Sayfa Sayısı
[dc.identifier.endpage]
159
Dergi Adı
[dc.relation.journal]
TEXAS HEART INSTITUTE JOURNAL
Dergi Sayısı
[dc.identifier.issue]
2
Dergi Cilt Bilgisi
[dc.identifier.volume]
33
Tek Biçim Adres
[dc.identifier.uri]
https://hdl.handle.net/20.500.12628/3365
Görüntülenme Sayısı ( Şehir )
Görüntülenme Sayısı ( Ülke )
Görüntülenme Sayısı ( Zaman Dağılımı )
Görüntülenme
39
09.12.2022 tarihinden bu yana
İndirme
1
09.12.2022 tarihinden bu yana
Son Erişim Tarihi
25 Şubat 2024 09:32
Google Kontrol
Tıklayınız
CarboMedics follow-up aortic months results (P=NS) Freedom events respectively Medical either patients mechanical thromboembolic deterioration structural compare bleeding obtained Definitive prospective randomized extensive trials judgment performance clinical excellent similar prostheses standard indicate long-term consecutive position
6698 sayılı Kişisel Verilerin Korunması Kanunu kapsamında yükümlülüklerimiz ve çerez politikamız hakkında bilgi sahibi olmak için alttaki bağlantıyı kullanabilirsiniz.

creativecommons
Bu site altında yer alan tüm kaynaklar Creative Commons Alıntı-GayriTicari-Türetilemez 4.0 Uluslararası Lisansı ile lisanslanmıştır.
Platforms